Product Features | |||||
Meet medical safety approvals | |||||
2 MOPP input to output isolation | |||||
Energy efficiercy Level VI | |||||
Touch current ≤ 100μA | |||||
≤0.5W standby power | |||||
Peak load:120% rated load(1s) | |||||
Up to 5000m operating altitude | |||||
Dial switch control output | |||||
Withless design featrues,natural air cooling |
Input
Input Voltage Range:90-264VAC(90-115VAC refer to derating curve)
Frequency Range:47-63Hz
Input Current:4.0A max at 100VAC
Inrush Current:120A max at 240VAC cold start
Touch Leakage Current(max):≤ 250μA at 264VAC
Environmental
Operationg Temperature: 0℃ to 30℃
Storage Temperature:- 25℃ to 70℃
Operating Humidity:5 to 90% RH,non-condensing
Storage Humidity::5 to 95% RH
Operating Altitude:5,000 M
General
Dimensions:165(L)×69.5(W)×24.25(H)mm
Weight:500g
MTBF:>100000hrs MIL-HKBK-217 at 25℃
EN60601
Safety Approvals
Safety Agency/Mark
Medical I.T.E.(Meet) CB IEC60601-1 IEC60950-1 IEC62368-1 UL
-
UL62368-1 TuV-SUD/Mark
- CCC
-
GB4943.1 CE
-
EN62368-1
Model Number | Voltage (V) | Current (A) | Rated Power (W) | Ripple & Noise (max) mVpk-pk | Voltage tolerance | Line& Load Regulation | Efficiency (Average) | Start Up Delay |
UES300DZ-190F70-SPA | 19.00 | 0-15.7 | 300.0 | 190 | ±5% | Line: ±1% | 92.0% | ≤3s |
UES300DZ-240C500-SPA | 24.00 | 0-12.5 | 300.0 | 240 | | 92.0% | ≤3s | |
UES300DZ-4800630-SPA | 48.00 | 0-6.3 | 300.0 | 480 | Load: ±5% | 92.0% | ≤3s |