Product Features | |||||
Medical safety approvals | |||||
2 MOPP input to output isolation | |||||
Energy efficiercy Level VI | |||||
Touch current ≤ 100μA | |||||
≤0.21W standby power | |||||
Peak load:200% rated load @ 230Vac | |||||
Up to 5000m operating altitude | |||||
Adjustable resistance to adjust output voltage | |||||
Withless design featrues,natural air cooling |
Input |
Input Voltage Range:90-264VAC(90-115VAC refer to derating curve)
Frequency Range:47-63Hz
Input Current:2.0A max at 100VAC
Inrush Current:120A max at 240VAC cold start
Touch Leakage Current(max):≤ 100μA at 264VAC
Environmental
Operating Temperature: -30℃ to 70℃(45℃ to 70℃ refer to derating curve)
Storage Temperature:- 40℃ to 80℃
Operating Humidity:5 to 90% RH,non-condensing
Storage Humidity::5 to 95% RH
Operating Altitude:5,000 M
General
Dimensions:101.6(L)×50.8(W)×29.0(H)mm
Weight:190g
MTBF:>200000hrs MIL-HKBK-217 at 25℃
Safety Approvals | |||||||||
Safety Agency/Mark | Medical | I.T.E. | |||||||
CB | IEC60601-1 | - | |||||||
TUV-SUD Mark | | EN60601-1 | | - | |||||
NRTL | | ANSI/AAMI ES60601-1 | | - |
Model Number | Voltage (V) | Current (A) | Rated Power (W) | Ripple & Noise (max) mVpk-pk | Voltage Tolerance | Line & Load Regulation | Efficiency | Start Up delay |
UES901-XXXYYYSPA-OP | 12.00 | 0.01-7.00 | 84 | 100 | ±5% | | 89% | ≤3s |
15.00 | 0.01-6.00 | 90 | 120 | Line: ±1% | 89% | ≤3s | ||
18.00 | 0.01-5.00 | 90 | 150 | 89% | ≤3s | |||
24.00 | 0.01-3.75 | 90 | 150 | Load: ±5% | 90% | ≤3s | ||
36.00 | 0.01-2.50 | 90 | 150 | 90% | ≤3s | |||
48.00 | 0.01-1.87 | 90 | 150 | 90% | ≤3s |