Product Features | |||||
Medical & ITE safety meet | |||||
2 MOPP input to output isolation | |||||
Energy efficiency Level V | |||||
Touch current ≤ 100μA | |||||
≤0.3W standby power | |||||
Up to 5000m operating altitude | |||||
The peak load can reach 7A and last for more than1s |
Input |
Input Voltage Range:90-264VAC
Frequency Range:47-63Hz
Input Current:1.5A max at 90VAC
Inrush Current:100A max at 240VAC cold start
Touch Leakage Current(max):≤ 100μA at 264VAC
Environmental
Operating Temperature: 0℃ to 40℃
Storage Temperature:- 20℃ to 70℃
Operating Humidity:10 to 90% RH,non-condensing
Storage Humidity::5 to 90% RH
Operating Altitude:5,000 M
General
Dimensions:134(L)×64(W)×32(H)mm
Weight:125g
MTBF:>100000hrs MIL-HKBK-217 at 25℃
Safety Approvals | |||||||||
Safety Agency/Mark | Medical | I.T.E.(Meet) | |||||||
CB | IEC60601-1 | IEC62368-1 |
Voltage (V) Current (A) Rated Power (W) Line: ±1% Load: ±5%
Model Number
Voltage Tolerance
Line & Load Regulation
Efficiency
Start Up delay
UE63-XXXYYYSPA-OP
18.0 0.01-3.50 63W 240 ±5%
87%
≤3s