Product Features | |||||
Medical safety approvals | |||||
2 MOPP input to output isolation | |||||
Energy efficiency Level VI | |||||
Touch current ≤ 100μA | |||||
≤0.15W standby power | |||||
Peak load:140W MAX @ Rated Input Voltage | |||||
Up to 5000m operating altitude | |||||
3 type of AC inlet |
Input |
Input Voltage Range:90-264VAC
Frequency Range:47-63Hz
Input Current:1.8A max at 90VAC
Inrush Current:120A max at 240VAC cold start
Touch Leakage Current(max):≤ 100μA at 264VAC
Environmental
Operating Temperature: 0℃ to 40℃
Storage Temperature:- 25℃ to 60℃
Operating Humidity:10 to 90% RH,non-condensing
Storage Humidity::5 to 90% RH
Operating Altitude:5,000 M
General
Dimensions:110.5(L)×56.5(W)×31.0(H)mm
Weight:260g
MTBF:>50000hrs MIL-HKBK-217 at 25℃
Safety Approvals | |||||||
Safety Agency/Mark | Medical(Meet) | ||||||
CB | IEC60601-1 | | |||||
UL | | ANSI/AAMI ES60601-1 CAN/CSA C22.2 NO.60601-1 | | ||||
TuV-SUD/Mark | | EN60601-1 | | |
Voltage (V) Current (A) Rated Power (W) Efficiency (Average) Line: ±1% Load: ±5% 13.1-14.0
Model Number
Voltage tolerance
Line& Load Regulation
Start Up Delay
UES65CZ-XXXYYYSPA 12.0 0.01-5.0 60 200 ±5%
88.0%
≤3s
12.1-13.0 0.01-4.61 60 200 ±5%
88.0%
≤3s
0.01-4.28 60 200 ±5%
88.0%
≤3s
14.1-15.0 0.01-4.00 60 200 ±5%
88.0%
≤3s
19.0 0.01-3.42 65 200 ±5%
88.0%
≤3s
19.1-20.0 0.01-3.25 65
240 ±5%
88.0%
≤3s
20.1-21.0 0.01-3.09 65
240 ±5%
88.0%
≤3s
21.1-22.0 0.01-2.95 65 240 ±5%
88.0%
≤3s
22.1-23.0 0.01-2.82 65
240 ±5%
88.0%
≤3s
23.1-24.0 0.01-2.71 65
240 ±5%
88.0%
≤3s