Product Features
Medical & I.T.E. safety approvals |
2 MOPP input to output isolation |
Touch current≤100μA |
Energy efficiency level VI |
≤0.075W standby power |
5V or 48V outputs,up to 36W |
Up to 5000m operating altitude |
Input |
Input Voltage Range:90-264VAC
Frequency Range:47-63Hz
Input Current:1.0A,at 90VAC
Inrush Current:70A max at 240VAC cold start
Touch Leakage Current(max):≤ 100μA at 264VAC
Environmental
Operating Temperature: 0℃ to 45℃
Storage Temperature:- 20℃ to 60℃
Operating Humidity:10 to 90% RH,non-condensing
Storage Humidity::5 to 90% RH
Operating Altitude:5,000 M
General
Dimensions:88.5(L)×31.8(W)×65.6(H)mm
Weight:210g
MTBF:>100000hrs MIL-HKBK-217 at 25℃
Safety Approvals | |||||||||
Safety Agency/Mark | Medical | I.T.E. | |||||||
CB | IEC60601-1 | | | | IEC62368-1 | | | | |
| ANSI/AAMI ES60601-1 | | | | | | | ||
UL | CAN/CSA-C22.2 NO.60601-1 | | | UL62368-1 | | | | ||
TuV SUD/Mark | EN60601-1 | | | | - | | | | |
TuV SUD/GS | - | | | EN62368-1 | |||||
CE | - | | | EN62368-1 | |||||
FCC | - | | | FCC PART 15 |
Voltage (V) Current (A) Rated Power (max) Ripple & Noise (Max) Voltage tolerance Line& Load Regulation Efficiency (Average) Start Up delay 14.1-12.0
Model Number
UES36LCP2-XXXYYYSPA
5.0 0.01-6.00 30.0 200 ±8%
85.00% ≤3s
5.1-5.9 0.01-5.00
30.0 200
±5%
88.30% ≤3s
7.5-14.0 0.01-4.00
36.0 200
±5%
88.30%
≤3s
0.01-2.40 36.0
200
±5%
88.30%
≤3s
21.1-30.0 0.01-1.63 36.0
200
±5%
88.30%
≤3s
30.1-48.0 0.01-1.16 36.0
300
±5%
88.30%
≤3s