Medical & I.T.E. safety approvals |
2 MOPP input to output isolation |
Touch current≤100μA |
≤0.075W standby power |
Up to 5000m operating altitude |
Input |
Input Voltage Range:90-264VAC
Frequency Range:47-63Hz
Input Current:0.5A,at 100VAC
Touch Leakage Current(max):≤ 100μA at 264VAC
Environmental
Operating Temperature: 0℃ to 45℃
Storage Temperature:- 20℃ to 60℃
Operating Humidity:10 to 90% RH,non-condensing
Storage Humidity::5 to 90% RH
Operating Altitude:5,000 M
General
Dimensions:47.5(L)×42.5(W)×27.5(H)mm
Weight:50g
MTBF:>100000hrs MIL-HKBK-217 at 25℃
Safety Approvals | |||||||||||
Safety Agency/Mark | Medical(Meet) | I.T.E.(Meet) | |||||||||
CB | IEC60601-1 | - | |||||||||
UL | ANSI / AAMI ES60601-1 | - | |||||||||
CAN / CSA-C22.2 NO.60601-1 | |||||||||||
TuV SUD/Mark | EN60601-1 | - | |||||||||
CCC | | | | - | | | | GB4943.1 | | | |
Model Number | Voltage (V) | Current (A) | Rated Power (W) | Ripple & Noise (max) mVpk-pk | Voltage tolerance | Line& Load Regulation | Efficiency (Average) | Start Up delay |
UES10LZ1-050200SPA UES10LZ2-050200SPA | 5.0 | 0.01-2.00 | 10 | 200 | ±5% | Line: ±5% Load: ±5% | 79% | ≤3s |