Product Features
Meets medical & I.T.E. safety |
2 MOPP input to output isolation |
Touch current≤10μA |
Energy efficiency level VI |
≤0.075W standby power |
4.2V to 12V outputs,up to 6W |
Up to 5000m operating altitude |
Interchangeable AC plugs |
Input |
Input Voltage Range:80-264VAC
Frequency Range:47-63Hz
Input Current:0.2A,at 90VAC
Inrush Current:30A max at 240VAC cold start
Touch Leakage Current(max):≤ 10μA at 264VAC
Environmental
Operationg Temperature: 0℃ to 45℃
Storage Temperature:- 20℃ to 60℃
Operating Humidity:10 to 90% RH,non-condensing
Storage Humidity::5 to 90% RH
Operating Altitude:5,000 M
General
Dimensions:74.0(L)×43.0(W)×33.0(H)mm
Weight:85g
MTBF:>100000hrs MIL-HKBK-217 at 25℃
TüV Rheinland-GS
Safety Approvals Safety Agency/Mark Medical I.T.E. CB IEC60601-1 / IEC60601-1-11 IEC62368-1 UL ANSI / AAMI ES60601-1 / 60601-1-11 UL62368-1 CAN / CSA-C22.2 NO。60601-1 TUV-Mark EN60601-1 / EN60601-1-11 - - EN62368-1 CE - EN62368 CCC - GB4943.1 PSE - J62368 KC - K60950-1 RCM - AS / NES62368
Models & Parameters
Model Number | Voltage | Current | Rated Power | Ripple & Noise | Voltage tolerance | Line& Load | Efficiency (Average) | Start Up delay |
(V) | (A) | (Max) | Regulation | |||||
UES06WNCP1-XXXYYYSPA | 4.2 | 0.01-1.20 | 5.04W | 200mVpk-pk | ±5% | Line: ±1% Load: ±5% | 73.83% | ≤3s |
5.0-6.0 | 0.01-1.00 | 6.00W | 150mVpk-pk | ±5% | 73.77% | ≤3s | ||
6.1-7.0 | 0.01-0.86 | 6.00W | 150mVpk-pk | ±5% | 78.16% | ≤3s | ||
7.1-8.0 | 0.01-0.75 | 6.00W | 200mVpk-pk | ±5% | 78.26% | ≤3s | ||
8.1-8.9 | 0.01-0.67 | 6.00W | 200mVpk-pk | ±5% | 78.38% | ≤3s | ||
9.0-10.0 | 0.01-0.60 | 6.00W | 200mVpk-pk | ±5% | 78.35% | ≤3s | ||
10.1-11.0 | 0.01-0.54 | 6.00W | 200mVpk-pk | ±5% | 78.42% | ≤3s | ||
11.1-12.0 | 0.01-0.50 | 6.00W | 200mVpk-pk | ±5% | 78.53% | ≤3s |